S0001307
Analyser, haematology, 18 parameter
Analyser, haematology, 18 parameter
Indicative Price USD 10,833.33
General Description
Analyser, haematology, 18 parameter
Product Description
Scope: Complete blood count (CBC), with 5-part differential, for
routine haematology
Platform: Bench-top, open, automated system
Method: Flow-cytometry, colorimetry, electrical impedance
Sample: Whole blood
Throughput: 50 to 100 samples per hour
Parameters:
Red blood cell (RBC), white blood cell (WBC),
haemoglobin (HGB), haematocrit (HCT), mean cell volume (MCV), Mean cell haemoglobin (MCH), red cell distribution (RDW-SD and RDW-CV),
platelets (PLT), platelet distribution (PDW-SD and PDW-CV), mean platelet volume (MPV), WBC differential (neutrophils, lymphocytes, monocytes, eosinophils and basophils)
Data management system built in for analysis and reporting, stores at least 100 patient data, including quality control and calibrators.
Results reportable via external printer.
Printed test results show sample ID, date and time, test results can be reported in graphs, histograms, scatter graphs
Quality control flags abnormal results, auto-calibration
Connects to external PC via USB, can interface to LIS (Laboratory Information System)
Casing is of corrosion-proof material
Power requirements: 220V/50Hz, with voltage surge protection
The equipment tolerates harsh environment, such as high humidity, operating temperature 15C to 35C
Supplied with
1 x set of standard accessories, consumables and parts required to operate the equipment:
1x power cable, 1x network cable, 1x barcode scanner, 1x CD-ROM of software, 1x diluent cap assembly, 1x support frame of the diluent cap, 1x lyse cap assembly, 1x waste liquid cap assembly and start-up kit (1x diluent, 1x lyse and 1x cleaner (UN1791, Class 8)
OBS Cleaning materials with hypochlorite solution are subject to dangerous goods classification (UN 1791, Class 8)
User manual in English; in French and Spanish upon request
Required, but not supplied
Training & installation
Reagents kits
Blood collection tubes with EDTA
UPS to ensure uninterrupted finalisation of on-going testing
PC, printer
Intended use
Used for haematology analysis for diagnosis
Used in hospital laboratories and reference centres
Split delivery of reagents is recommended due to their limited shelf life
Manufacturers recommended waste disposal:
Packaging and Labelling
Primary packaging:
Unit presentation: 1(one) Analyser, haematology,18 parameter
Weight - Volume
Estimated weight: 90kg
Estimated volume: 0.201m3